SPECIFICATION
MODEL
DC VOLTAGE
RPS-120S-15
RPS-120S-24
RPS-120S-27
RPS-120S-48
12V
15V
24V
27V
48V
Peak(10 sec.) 11.8A
9.5A
6.25A
5.55A
3.125A
Convection 9.5A
7.6A
5A
4.44A
2.5A
Peak(10 sec.) 141.6W
142.5W
150W
149.8W
150W
Convection 114W
114W
120W
119.9W
120W
RIPPLE & NOISE (max.) Note.2 100mVp-p
120mVp-p
150mVp-p
150mVp-p
200mVp-p
VOLTAGE ADJ. RANGE
CURRENT
RATED
POWER
OUTPUT
RPS-120S-12
14.3~15.8V
22.8~25.2V
25.6 ~ 28.4V
45.6 ~50.4V
VOLTAGE TOLERANCE Note.3 ±2.0%
±2.0%
±1.0%
±1.0%
±1.0%
±0.5%
±0.5%
±0.5%
±0.5%
±0.5%
LOAD REGULATION
±1.0%
±1.0%
±1.0%
±1.0%
±1.0%
SETUP, RISE TIME
600ms, 30ms/230VAC
HOLD UP TIME (Typ.)
15ms/230VAC
94%
93.5%
LINE REGULATION
11.4~12.6V
15ms/115VAC at full load
VOLTAGE RANGE Note.4 80 ~ 264VAC
INPUT
113 ~ 370VDC
FREQUENCY RANGE
47 ~ 63Hz
POWER FACTOR
PF>0.94/230VAC
EFFICIENCY (Typ.)
91%
AC CURRENT (Typ.)
2.3A/115VAC
PF>0.98/115VAC at full load
92%
93%
1.1A/230VAC
INRUSH CURRENT (Typ.)
COLD START
LEAKAGE CURRENT(max.) Note.5
Earth leakage current < 150μA/264VAC , touch current < 80μA/264VAC
OVERLOAD
PROTECTION
600ms, 30ms/115VAC at full load
30A/115VAC
130~160% rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is removed
16.5 ~ 19.5V
13.2 ~ 15.6V
OVER VOLTAGE
60A/230VAC
26.4 ~ 31.2V
29.7 ~ 35V
52.8 ~ 62.4V
Protection type : Shut down o/p voltage, re-power on to recover
OVER TEMPERATURE
Protection type : Shut down o/p voltage, recovers automatically after temperature goes down
WORKING TEMP.
-30 ~ +85℃ (Refer to "Derating Curve")
WORKING HUMIDITY
20 ~ 90% RH non-condensing
ENVIRONMENT STORAGE TEMP.
-40 ~ +85℃
TEMP. COEFFICIENT
±0.03%/℃ (0 ~ 50℃)
VIBRATION
10 ~ 500Hz, 2G 10min./1cycle, 60min. each along X, Y, Z axes
OPERATING ALTITUDE Note.6 4000 meters
SAFETY STANDARDS
IEC60601-1, TUV BS EN/EN60601-1, EAC TP TC 004 , UL ANSI / AAMI ES60601-1 (3.1 version),
CAN/CSA-C22.2 No. 60601-1:14 - Edition 3 approved; Design refer to BS EN/EN60335-1 (By request)
ISOLATION RESISTANCE Primary-Secondary: 2xMOPP, Primary-Earth:1xMOPP, Secondary-Earth:1xMOPP
WITHSTAND VOLTAGE
I/P-O/P:4KVAC
I/P-FG:2KVAC
O/P-FG:1.5KVAC
ISOLATION RESISTANCE I/P-O/P, I/P-FG, O/P-FG: 100M Ohms / 500VDC / 25℃/ 70% RH
EMC EMISSION
SAFETY &
EMC
(Note 7)
EMC IMMUNITY
OTHERS
NOTE
Parameter
Standard
Conducted emission
BS EN/EN55011 (CISPR11)
Class B
Radiated emission
BS EN/EN55011 (CISPR11)
ClassⅠ: Class B , Class Ⅱ: Class A
Harmonic current
Voltage flicker
BS EN/EN60601-1-2
BS EN/EN61000-3-2
Class A
BS EN/EN61000-3-3
-----
Parameter
Standard
Test Level / Note
ESD
BS EN/EN61000-4-2
Level 4, 15KV air ; Level 4, 8KV contact
RF field susceptibility
BS EN/EN61000-4-3
Level 3, 10V/m( 80MHz~2.7GHz )
Table 9, 9~28V/m( 385MHz~5.78GHz )
EFT bursts
Surge susceptibility
Conducted susceptibility
BS EN/EN61000-4-4
Level 3, 2KV
BS EN/EN61000-4-5
Level 4, 4KV/Line-FG; 2KV/Line-Line
BS EN/EN61000-4-6
Level 3, 10V
Magnetic field immunity
BS EN/EN61000-4-8
Voltage dip, interruption
BS EN/EN61000-4-11
Level 4, 30A/m
95% dip 1 periods, 30% dip 25 periods,
95% interruptions 250 periods
MTBF
4050.3K hrs min.
DIMENSION (L*W*H)
76.2*50.8*28mm or 3" * 2" *1.1" inch
PACKING
0.13Kg; 100pcs/14Kg/1.13CUFT
Telcordia SR-332 (Bellcore) ; 468.0K hrs min.
Test Level / Note
MIL-HDBK-217F (25℃)
μ
μ
6. ℃℃
(as available on http://www.meanwell.com)
※:https://www.meanwell.com/serviceDisclaimer.aspx
File Name:RPS-120S-SPEC 2022-03-18
120W 3"×2" Reliable Green Medical Power Supply
R P S - 1 2 0 S series
Block Diagram
fosc : 85KHz
RECTIFIERS
&
PFC
EMI
FILTER
I/P
POWER
SWITCHING
RECTIFIERS
&
FILTER
+V
-V
FG
O.T.P.
DETECTION
CIRCUIT
O.L.P.
CONTROL
O.V.P.
Derating Curve
Convection
100
80
70
60
LOAD (%)
50
30
-30
0
10
20
30
40
50
70
85 (HORIZONTAL)
AMBIENT TEMPERATURE (℃)
Output Derating VS Input Voltage
(for 24V/27V/48V)
120W
114W
(for 12V/15V)
100W
LOAD (W)
80W
80
90
110
120
130
140
160
180
200
220
240
264
INPUT VOLTAGE (VAC) 60Hz
File Name:RPS-120S-SPEC 2022-03-18
120W 3"×2" Reliable Green Medical Power Supply
R P S - 1 2 0 S series
Mechanical Specification
3.175
Unit:mm
50.8
FS1
T1
C5
44.45
CN1
T3.15A/250V
SVR1
HS1
C100
CN100
1
2
3
4
3
2
1
FS2 T3.15A
/250V
C105
C1
C119
4 -ψ
3.3
69.9
3.175
3 max.
28
76.2
AC Input Connector (CN1) : JST B3P-VH or equivalent
Pin No.
1
2
3
Assignment
AC/L
No Pin
AC/N
Mating Housing
Terminal
JST VHR
or equivalent
JST SVH-21T-P1.1
or equivalent
DC Output Connector (CN2) : JST B4P-VH or equivalent
Pin No.
1,2
3,4
Assignment
+V
-V
Mating Housing
Terminal
JST VHR
or equivalent
JST SVH-21T-P1.1
or equivalent
1.HS1 must have safety isolation distance
with system case.
※Note :
1.RPS-120S model delivers EMI Class B for both conducted emission and radiated emission for the power supply,
when configured into Class Ⅰ(with FG) system.
2.RPS-120S model delivers EMI Class B conducted emission and Class A radiated emission with King Core K5B RC (12*15*7)
in output cable for the power supply when configured into Class Ⅱ (no FG) system.
Installation Manual
Please refer to : http://www.meanwell.com/manual.html
File Name:RPS-120S-SPEC 2022-03-18
很抱歉,暂时无法提供与“RPS-120S-24”相匹配的价格&库存,您可以联系我们找货
免费人工找货- 国内价格
- 1+257.25302
- 10+246.96415
- 50+236.66486