0
登录后你可以
  • 下载海量资料
  • 学习在线课程
  • 观看技术视频
  • 写文章/发帖/加入社区
创作中心
发布
  • 发文章

  • 发资料

  • 发帖

  • 提问

  • 发视频

创作活动
100

100

  • 厂商:

    TPI(TPI 公司)

  • 封装:

    -

  • 描述:

    CLIP BATTERY AA CELL INSUL STEEL

  • 数据手册
  • 价格&库存
100 数据手册
100 EZ Digital Multimeter Instruction Manual SAFETY CONSIDERATIONS WARNING: Please follow manufacturers test procedures whenever possible. Do not attempt to measure unknown voltages or components until a complete understanding of the circuit is obtained. GENERAL GUIDELINES ALWAYS • • • • • • Test the 100 before using it to make sure it is operating properly. Inspect the test leads before using to make sure there are no breaks or shorts. Double check all connections before testing. Have someone check on you periodically if working alone. Have a complete understanding of the circuit being measured. Disconnect power to circuit, then connect test leads to the 100, then to circuit being measured. CATEGORY RATINGS DEFINITIONS IEC 1010 Over Voltage: Cat II - 1000V CAT III - 600V Pollution Degree 2 CAT II - 1000V Installation Category (Overvoltage Category) II: Includes voltages encountered on the step down side of the transformer on the building and at a distance of 10 meters from the CAT III source. CAT III - 600V Installation Category (Overvoltage Category III: Includes voltage encountered on the distribution level with short distance to the main service connection. Pollution Degree 2 Normally only nonconductive pollution occurs. Occasionally temporary conductivity caused by condensation must be expected. NEVER • • • Attempt to measure unknown high voltages. Connect the test leads to a live circuit before setting up the instrument. Touch any exposed metal part of the test lead assembly. 2 3 SAFETY CONSIDERATIONS WARNING: Please follow manufacturers test procedures whenever possible. Do not attempt to measure unknown voltages or components until a complete understanding of the circuit is obtained. GENERAL GUIDELINES ALWAYS • • • • • • Test the 100 before using it to make sure it is operating properly. Inspect the test leads before using to make sure there are no breaks or shorts. Double check all connections before testing. Have someone check on you periodically if working alone. Have a complete understanding of the circuit being measured. Disconnect power to circuit, then connect test leads to the 100, then to circuit being measured. CATEGORY RATINGS DEFINITIONS IEC 1010 Over Voltage: Cat II - 1000V CAT III - 600V Pollution Degree 2 CAT II - 1000V Installation Category (Overvoltage Category) II: Includes voltages encountered on the step down side of the transformer on the building and at a distance of 10 meters from the CAT III source. CAT III - 600V Installation Category (Overvoltage Category III: Includes voltage encountered on the distribution level with short distance to the main service connection. Pollution Degree 2 Normally only nonconductive pollution occurs. Occasionally temporary conductivity caused by condensation must be expected. NEVER • • • Attempt to measure unknown high voltages. Connect the test leads to a live circuit before setting up the instrument. Touch any exposed metal part of the test lead assembly. 2 3 MEASURING AC/DC VOLTAGE SPECIFICATIONS Function Range Resolution DCV 1.5V- 4V 40.00V 400.0V 600V 1.7V-4V 40.00V 400.0V 600V 400.0 4.000k 40.00k 400.0k 4.000M 40.00M 0.001V 0.01V 0.1V 0.1V 0.001V 0.01V 0.1V 1V 0.1 0.001k 0.01k 0.1k 0.001M 0.01M ACV OHM Accuracy ±(0.5% + 4) ±(0.8% + 4) ±(0.8% + 4) ±(1.2% + 4) ±(0.8% + 4) ±(2.0% + 4) GENERAL SPECIFICATIONS Power Supply Battery Life Size (H x L x W) Weight 2 Each 1.5 Volt “AA” Batteries 560 hrs. Alkaline 45mm x 78mm x 153mm (1.8” x 3.1” x 6.0”) 340g (12 oz) 4 Make sure that the ground and positive leads are plugged into the proper receptacle for corresponding function positions. Do not attempt to make a voltage measurement of more than 500V or of a voltage that is unknown. Measurement Procedure: 1. Disconnect power to circuit to be measured. 2. Plug black test lead into the COM input jack. 3. Plug the red test lead into the V/W input jack. 4. Set the rotary switch to the “ON” position. 5. Connect test leads to circuit to be measured. 6. Reconnect power to circuit to be measured. 7. Read the voltage on the TPI 100. Note: For Auto mode to operate properly, DC voltage must be between +0.7 and 600V. AC voltage must be between 3 and 500V. 5 MEASURING AC/DC VOLTAGE SPECIFICATIONS Function Range Resolution DCV 1.5V- 4V 40.00V 400.0V 600V 1.7V-4V 40.00V 400.0V 600V 400.0 4.000k 40.00k 400.0k 4.000M 40.00M 0.001V 0.01V 0.1V 0.1V 0.001V 0.01V 0.1V 1V 0.1 0.001k 0.01k 0.1k 0.001M 0.01M ACV OHM Accuracy ±(0.5% + 4) ±(0.8% + 4) ±(0.8% + 4) ±(1.2% + 4) ±(0.8% + 4) ±(2.0% + 4) GENERAL SPECIFICATIONS Power Supply Battery Life Size (H x L x W) Weight 2 Each 1.5 Volt “AA” Batteries 560 hrs. Alkaline 45mm x 78mm x 153mm (1.8” x 3.1” x 6.0”) 340g (12 oz) 4 Make sure that the ground and positive leads are plugged into the proper receptacle for corresponding function positions. Do not attempt to make a voltage measurement of more than 500V or of a voltage that is unknown. Measurement Procedure: 1. Disconnect power to circuit to be measured. 2. Plug black test lead into the COM input jack. 3. Plug the red test lead into the V/W input jack. 4. Set the rotary switch to the “ON” position. 5. Connect test leads to circuit to be measured. 6. Reconnect power to circuit to be measured. 7. Read the voltage on the TPI 100. Note: For Auto mode to operate properly, DC voltage must be between +0.7 and 600V. AC voltage must be between 3 and 500V. 5 MEASURING RESISTANCE Do not attempt to make resistance measurements with circuit energized. For best results, remove the resistor completely from the circuit before measuring. NOTE: To make accurate low ohm measurements, short the test leads together and record the resistance reading. Deduct this value from actual readings. Measurement Procedure: 1. Disconnect power to circuit to be measured. 2. Plug black test lead into the COM input jack. 3. Plug the red test lead into the V/W input jack. 4. Set the rotary switch to the “ON” position. 5. Connect test leads to circuit to be measured. 7. Read the resistance on the TPI 100. 6 MAINTENANCE 1. Battery Replacement: The 265 will display a battery symbol when the internal 9 Volt battery needs replacement. The battery is replaced as follows: a. Disconnect and remove all test leads from live circuits and from the 265. b. Loosen the screw from the back of the 265 battery cover. c. Remove the battery compartment cover. d. Remove old battery and replace with new battery. Observe the correct polarity on the battery. e. Reassemble the instrument in reverse order from above. 2. Cleaning your 265: Use a mild detergent and slightly damp cloth to clean the surfaces of the 265. 7 MEASURING RESISTANCE Do not attempt to make resistance measurements with circuit energized. For best results, remove the resistor completely from the circuit before measuring. NOTE: To make accurate low ohm measurements, short the test leads together and record the resistance reading. Deduct this value from actual readings. Measurement Procedure: 1. Disconnect power to circuit to be measured. 2. Plug black test lead into the COM input jack. 3. Plug the red test lead into the V/W input jack. 4. Set the rotary switch to the “ON” position. 5. Connect test leads to circuit to be measured. 7. Read the resistance on the TPI 100. 6 MAINTENANCE 1. Battery Replacement: The 265 will display a battery symbol when the internal 9 Volt battery needs replacement. The battery is replaced as follows: a. Disconnect and remove all test leads from live circuits and from the 265. b. Loosen the screw from the back of the 265 battery cover. c. Remove the battery compartment cover. d. Remove old battery and replace with new battery. Observe the correct polarity on the battery. e. Reassemble the instrument in reverse order from above. 2. Cleaning your 265: Use a mild detergent and slightly damp cloth to clean the surfaces of the 265. 7 SAFETY CONSIDERATIONS WARNING: Please follow manufacturers test procedures whenever possible. Do not attempt to measure unknown voltages or components until a complete understanding of the circuit is obtained. GENERAL GUIDELINES ALWAYS • • • • • • Test the 100 before using it to make sure it is operating properly. Inspect the test leads before using to make sure there are no breaks or shorts. Double check all connections before testing. Have someone check on you periodically if working alone. Have a complete understanding of the circuit being measured. Disconnect power to circuit, then connect test leads to the 100, then to circuit being measured. CATEGORY RATINGS DEFINITIONS IEC 1010 Over Voltage: Cat II - 1000V CAT III - 600V Pollution Degree 2 CAT II - 1000V Installation Category (Overvoltage Category) II: Includes voltages encountered on the step down side of the transformer on the building and at a distance of 10 meters from the CAT III source. CAT III - 600V Installation Category (Overvoltage Category III: Includes voltage encountered on the distribution level with short distance to the main service connection. Pollution Degree 2 Normally only nonconductive pollution occurs. Occasionally temporary conductivity caused by condensation must be expected. NEVER • • • Attempt to measure unknown high voltages. Connect the test leads to a live circuit before setting up the instrument. Touch any exposed metal part of the test lead assembly. 2 3 MEASURING AC/DC VOLTAGE SPECIFICATIONS Function Range Resolution DCV 1.5V- 4V 40.00V 400.0V 600V 1.7V-4V 40.00V 400.0V 600V 400.0 4.000k 40.00k 400.0k 4.000M 40.00M 0.001V 0.01V 0.1V 0.1V 0.001V 0.01V 0.1V 1V 0.1 0.001k 0.01k 0.1k 0.001M 0.01M ACV OHM Accuracy ±(0.5% + 4) ±(0.8% + 4) ±(0.8% + 4) ±(1.2% + 4) ±(0.8% + 4) ±(2.0% + 4) GENERAL SPECIFICATIONS Power Supply Battery Life Size (H x L x W) Weight 2 Each 1.5 Volt “AA” Batteries 560 hrs. Alkaline 45mm x 78mm x 153mm (1.8” x 3.1” x 6.0”) 340g (12 oz) 4 Make sure that the ground and positive leads are plugged into the proper receptacle for corresponding function positions. Do not attempt to make a voltage measurement of more than 500V or of a voltage that is unknown. Measurement Procedure: 1. Disconnect power to circuit to be measured. 2. Plug black test lead into the COM input jack. 3. Plug the red test lead into the V/W input jack. 4. Set the rotary switch to the “ON” position. 5. Connect test leads to circuit to be measured. 6. Reconnect power to circuit to be measured. 7. Read the voltage on the TPI 100. Note: For Auto mode to operate properly, DC voltage must be between +0.7 and 600V. AC voltage must be between 3 and 500V. 5 MEASURING RESISTANCE Do not attempt to make resistance measurements with circuit energized. For best results, remove the resistor completely from the circuit before measuring. NOTE: To make accurate low ohm measurements, short the test leads together and record the resistance reading. Deduct this value from actual readings. Measurement Procedure: 1. Disconnect power to circuit to be measured. 2. Plug black test lead into the COM input jack. 3. Plug the red test lead into the V/W input jack. 4. Set the rotary switch to the “ON” position. 5. Connect test leads to circuit to be measured. 7. Read the resistance on the TPI 100. 6 MAINTENANCE 1. Battery Replacement: The 265 will display a battery symbol when the internal 9 Volt battery needs replacement. The battery is replaced as follows: a. Disconnect and remove all test leads from live circuits and from the 265. b. Loosen the screw from the back of the 265 battery cover. c. Remove the battery compartment cover. d. Remove old battery and replace with new battery. Observe the correct polarity on the battery. e. Reassemble the instrument in reverse order from above. 2. Cleaning your 265: Use a mild detergent and slightly damp cloth to clean the surfaces of the 265. 7 The Value Leader TM www.tpi-thevalueleader.com 100 Instruction Manual Test Products International, Inc. 9615 SW Allen Blvd Beaverton, OR 97005-4814 USA Ph: 503-520-9197 Fax: 503-520-1225 Test Product International Ltd. 342 Bronte Street South, Unit 9 Milton, Ontario L9T 5B7 Ph: 905-693-8558 Fax: 905-693-0888 Test Products International Europe Longley House, International Drive, Southgate Crawley, West Sussex RH10 6AQ Ph: +44 (0) 1293 561212 Fax: +44 (0) 1293 813465 Copyright © 2006, Test Products International, Inc

很抱歉,暂时无法提供与“100”相匹配的价格&库存,您可以联系我们找货

免费人工找货