INSTRUCTION MANUAL
Models: H100
Light Meter
Table of Content
1. Product Introduction ………………………………………… 3
1-1 Features ………………….………………………… 3
1-2 Applications ……………………………………...… 3
2. Safety Information …………………………………………… 4
2-1 Cautions ………………………………..………..… 4
2-2 Safety Symbols ………………………..…..……… 4
3. Specifications ……………………………………...…………. 5
4. Operation Instructions ………………………………………
4-1 Unit Diagram ………………………………………
4-2 Measuring Procedure ……………………….…...
4-3 Zero Adjust ………………………..………………
4-4 Data Hold ……………………………….…………
4-5 Peak Hold ……………………....…………………
4-6 Data Record ………………………………………
7
7
8
8
8
8
9
5. Auto Power Off Disable ……………………………………
9
6. RS232 PC Serial Interface Information ………………….
10
7. Battery Replacement ..……………………………………..
11
8. Service Information …………………………………..…….
12
9. Warranty Information ………………………………………. 13
To see all available accessories and downloadable software, please visit
www.anaheimscientific.com
1. Product Introduction
1-1 Features
Two year warranty
Large LCD display with bar graph
Wide measurement ranges: 40.00/400.0/4,000/40,000/400,000lux
Display resolution: 0.04lux to 100lux, 0.01ft-cd to 10ft-cd
lux or foot-candle unit selection
4 light type select (Tungsten, Fluorescent, Sodium or Mercury)
Sensor meet C.I.E. spectrum, 2 filters
Data hold, Record (max., min.)
DC 1.5V battery ( UM-4, AAA) x 6
Longer battery life compared to 9V
1-2 Applications
Science Experiments
Photography
Manufacturing
Agriculture
Artistry
3
2. Safety Information
Read the following safety information carefully before attempting to operate or
service the meter. Only qualified personnel should perform repairs or servicing not
covered in this manual.
2-1 Cautions!
DO NOT submerge the products mentioned in this manual in water or any other
types of liquids.
This product is not designed for use in medical applications. The product can only
be used to measure body temperature simply for reference.
2-2 Safety Symbols
Dangerous, refer to this manual before using the meter.
Conforms to requirements of European Union and European Fare Trade
Association (EFTA).
Battery level is low.
Don’t dispose this product as unsorted municipal waste.
This instrument conforms to the following standards:
EN61326: Electrical equipment for measurement, control and laboratory use.
IEC61000-4-2: Electrostatic discharge immunity test.
IEC61000-4-3: Radiated, radio-frequency, electromagnetic field immunity test.
IEC61000-4-8: Power frequency magnetic field immunity test.
4
3. Specifications
Display
Measurement Ranges
Unit of Measurement
Measurable Lighting Sources
Sensor
Display values:
Unit
lux
fc
(foot-candle)
Accuracies:
lux
fc
(foot-candle)
Zero Adjust
Peak Hold
Data Hold
Memory
Power Off
Over and Under Range
Indicator
5 Digit LCD with bar graph indicator
40.00/400.0/4,000/40,000/400,000lux
lux or Foot-Candle (fc)
Tungsten, Fluorescent, Sodium or Mercury Lamps
Photo diode & color correction filter, spectrum meets
C.I.E. Cosine correction factor standards
Range
40
400
4,000
40,000
400,000
4
40
400
4,000
40,000
Range
40
400
4,000
40,000
400,000
4
40
400
4,000
40,000
In Range Display
0 to 40.00 Lux
36.0 to 400.0 Lux
360 to 4,000 Lux
3,600 to 40,000 Lux
10,000 to 400,000 Lux
0 to 3.720 Fc
3.35 to 37.20 Fc
33.5 to 372.0 Fc
335 to 3,720 Fc
930 to 37,200 Fc
Resolution
0.01 Lux
0. 1 Lux
1 Lux
10 Lux
100 Lux
0.001 Fc
0.01 Fc
0. 1 Fc
1 Fc
10 Fc
Accuracy**
± ( 3% rdg + 0.5 % F.S. )
< 100,000 Lux : ± ( 3% rdg + 0.5 % F.S. )*
± ( 3% rdg + 0.5 % F.S. )
< 9,300 Fc : ± ( 3% rdg + 0.5 % F.S. )*
40.00lux range only
Displays peak value
Display freeze
Saves MAX. or MIN values with recall
Auto or manual control
Over range indicator: " _ _ _ _ " (top of the display)
Under range indicator: "_ _ _ _" (bottom of the display)
5
Data Output
Operating Environment
Power Supply
RS-232 Serial data output
32° to 122°F (0° to 50°C),